Characterization of AlF3-passivated aluminum
mirrors using non-contact thermal metrology

Research summary · How SSTR maps the chemistry and defects of ultrathin mirror coatings faster and at finer resolution than XPS or ellipsometry

The problem

Aluminum mirrors for far-ultraviolet space optics are protected by a fluoride layer only about 16 to 25 nm thick. It’s invisible to the eye, but if it’s missing or uneven, the mirror loses reflectance below 200 nm. Checking that layer across a large reflector is slow. XPS needs a vacuum chamber, takes minutes per point and only sees the top ~10 nm. Ellipsometry uses a millimeter-scale spot. ATR-FTIR needs physical contact with a delicate optical surface. Raman couldn’t detect the layer at all.

The approach

A team from Laser Thermal, the U.S. Naval Research Laboratory and NASA Goddard Space Flight Center used steady-state thermoreflectance (SSTR) to inspect AlF₃-passivated aluminum mirrors. Because SSTR measures the sample’s thermal response, not just its optical reflectance, it doesn’t need the optical properties of the coating in advance. The team used Laser Thermal’s fiber-based SSTR-F system, with automated alignment and a motorized stage, to measure single points and raster-scan areas of several square millimeters.

To test sensitivity, they made three samples with physical masks: one half coated, one with a pattern of ~230 µm coated dots, and one with a gradual coating gradient.

What the research showed

TechniqueSpot sizeTime per pointDetects the AlF3 layer
SSTR1–10 µm~1 syes
Ellipsometry0.3 × 1.2 mm~5 syes
XPS~200 µm>1 minyes
ATF-FTIR (contact)~0.5 mm~30 syes
RamanDiffraction limited>10 sno

Why it means for your work

If you need to…This paper shows…
Check coating uniformity across a large opticSSTR raster-maps millimeter areas at micrometer resolution
Detect chemical changes, not just scratchesThermoreflectance separates chemistry from topography
Avoid contact or vacuumSSTR is optical, non-contact and runs in air
Inspect coatings with unknown optical constantsThe thermal measurement doesn’t need them

From paper to platform

SSTR-F was the earlier name of the instrument now called FASTR. FASTR’s SSTR mode runs on the same fiber-based platform with automated alignment and thermal property imaging.

Andrew H. Jones, John T. Gaskins, Patrick E. Hopkins, Scott G. Walton, David R. Boris, John P. Murphy, Luis V. Rodriguez de Marcos, Javier Del Hoyo, and Manuel A. Quijada, “Characterization of AlF₃-passivated aluminum mirrors using non-contact thermal metrology,” Review of Scientific Instruments 96, 024901 (2025). DOI: 10.1063/5.0241547. Open access under CC BY-NC-ND 4.0.